34 Installation and Specifications Monitor Safety Specifications
340
Measurement Server M3001A, Measurement Server Extensions M3016A M3015A, M3014A,
M3012A, Measurement Modules, and Flexible Module Server M8048A
Item Condition Range
Temperature Range Operating 0 to 45 °C (32 to 113 °F)
Non-operating
-40 to 70 °C (-40 to 158 °F)
Humidity Range Operating 95% Relative Humidity (RH) max. @ 40
°C (104 °F). M3015A
only non-condensing.
Non-operating
90% Relative Humidity (RH) max. @ 65 °C (150 °F)
Altitude Range Operating -500 m to 4600 m (-1600 to 15000 ft)
Non-operating
-500 m to 15300 m (-1600 to 50000 ft)
D80 Intelligent Display
Item Condition Range
Temperature Range Operating 0 to 35 °C (32 to 95 °F)
Non-operating
-20 to 60 °C (-4 to 140 °F)
Battery storage
-20 to 50 °C (-4 to 122 °F)
Humidity Range Operating 20% to 85% Relative Humidity (RH) (non condensing)
Non-operating
5% to 85% Relative Humidity (RH)
Altitude Range Operating -500 m to 3000 m (10000 ft)
Non-operating
-500 m to 12000 m (40000 ft)
Docking Station M8045A
Item Condition Range
Temperature Range Operating 0 to 40 °C (32 to 100 °F)
Non-operating
-20 to 60 °C (-4 to 140 °F)
Humidity Range Operating 20% to 85% Relative Humidity (RH) (non condensing)
Non-operating
5% to 85% Relative Humidity (RH)
Altitude Range Operating -500 m to 3000 m (10000 ft)
Non-operating
-500 m to 12000 m (40000 ft)
Thermal Array Recorder Module M1116B
Item Condition Range
Temperature Range Operating +5 to 45 °C (41 to 113°F)
Non-operating
-10 to 70 °C (14 to 158 °F)