Philips MP40/50 Medical Alarms User Manual


 
1 Basic Operation Related Products
12
WARNING The measurement server extensions can only function when they are connected to a measurement
server. If the measurement server is removed during monitoring, the measurements from both the
measurement server and the measurement server extension are lost.
M3014A, M3015A and M3016A Capnography Extensions
The M3014A Capnography Extension adds mainstream capnography, and optionally one pressure
plus either a pressure or a temperature, Cardiac Output and Continuous Cardiac Output to the MMS.
The optional M3015A Microstream CO
2
Extension adds microstream capnography and optionally
either pressure or temperature to the MMS. The optional M3016A Mainstream CO
2
Extension adds
mainstream capnography and optionally either pressure or temperature to the MMS.
1
Pressure connectors (red)
5
Inlet
2
Temperature connector (brown)
6
Microstream
connector CO
2
3
Mainstream connector CO
2
(optional)
7
Gas sample outlet
4
Cardiac Output connector
M3014A Capnography
M3015A Microstream
1
2
6
1
3
2
7
5
4
1
2
M3016A Mainstream