Philips MP40/50 Medical Alarms User Manual


 
35 Default Settings Appendix Invasive Pressure Default Settings
376
Invasive Pressure Default Settings
ABP, ART, Ao, BAP,
FAP, P, P1, P2, P3,
P4, UAP Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Alarms from Sys.
High Limit 160/ 90 (110) 120/ 70 (90) 90/ 60 (70) 180/ 90 (110)
Low Limit 90/ 50 (70) 70/ 40 (50) 55/ 20 (35) 70/ 50 (70)
Alarms On
Scale 150 100 100
Speed 25mm/s
Mean Only No
Filter 12 Hz
Mercury Cal. Yes
Artifact Suppr. 60 sec
Unit mmHg
Color Red
CV P, RA P, LA P,
UVP Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Alarms from Mean
High Limit 14 / 6 (10) 10 / 2 (4) 10 / 2 (4)
Low Limit 6 / -4 (0) 2 / -4 (0) 2 / -4 (0)
Alarms On
Scale 30
Speed 25mm/s
Mean Only Yes
Filter 12 Hz
Mercury Cal. Yes
Artifact Suppr. 60 sec
Unit mmHg
Color Cyan Blue
Invasive Pressure
Settings
PAP Settings ICP, IC1, IC2 Settings
Factory Adult Factory Pedi Factory Neo Factory Adult Factory Pedi Factory Neo
Alarms from Dia. Mean
High Limit 35 / 16 (20) 60 / 4 (26) 60 / 4 (26) 14 / 6 (10) 10 / 2 (4) 10 / 2 (4)
Low Limit 10 / 0 (0) 24 / -4 (12) 24 / -4 (12) 6 / -4 (0) 2 / -4 (0) 2 / -4 (0)
Alarms On On
Scale 30 30
Speed 25mm/s 25mm/s