Philips MP40/50 Medical Alarms User Manual


 
35 Default Settings Appendix ECG, Arrhythmia, and ST Default Settings
372
Arrhythmia Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Arrhythmia On Off Off
Pause Threshold 2.0 sec 1.5 sec
VTach HR 100 120 150
VTach Run 5
Vent Rhythm 14
SVT HR 180 200 210
SVT Run 5
PVCs/min 10 5 5
Non-Sustain On
Vent Rhythm On
Run PVCs On
Pair PVCs On
Off
R-On-T PVCs On
Off
V. Bi g emi n y On
Off
V. Tri g emi n y On
Off
PVCs/min On
Off
Multif. PVCs On
Off
Pacer N. Cap On
Pacer N. Pac On
Pause On
Off
Missed Beat On
Off
SVT On
IrregularHR On
Off
TimeOut 1st 3 min
TimeOut 2nd 10 min
Arrhy Off Message Yes
SOME ECG IN On
Off
Lead-independent ST Settings Factory Adult Factory Pedi Factory Neo
ST Alarm Mode Single ST
Alarms On
ST Analysis On Off
ST-Index On
ISO Point -80 ms
J Point 48 ms
ST Point J+60