Philips MP40/50 Medical Alarms User Manual


 
Measurement Specifications 34 Installation and Specifications
363
SvO
2
EEG
Complies with IEC 60601-2-26:1994/EN60601-2-26:1994.
tcpCO
2
response time, including transducer < 50 s
Warm-up Time < 3 minutes
Site Timer 0.5, 1, 1.5, 2, 2.5, 3, 4, 5, 6, 7, or 8 hours. Change Site alarm
when site time is expired and configurable automatic heating
switch-off
Transducer Heating Available Temperatures -
37.0°C, 41.0 to 45°C in steps of 0.5°C
tcGas Alarm Specifications Range Adjustment Alarm Delay
tcpO
2
10 to 745 mmHg
1.0 to 99.5 kPa
10 to 30 mmHg: 1 mmHg
1.0 to 4.0 kPa: 0.1 kPa
32 to 100 mmHg: 2 mmHg
4.2 to 13 kPa: 0.2 kPa
105 to 745mmHg: 5 mmHg
13.5 to 99.5 kPa: 0.5 kPa
10 seconds after the
value exceeds the set
limit range.
tcpCO
2
10 to 195 mmHg
1.0 to 26 kPa
10 to 30 mmHg: 1 mmHg
1.0 to 4.0 kPa: 0.1 kPa
32 to 100 mmHg: 2 mmHg
4.2 to 13 kPa: 0.2 kPa
105 to 195 mmHg: 5 mmHg
13.5 to 26 kPa: 0.5 kPa
tcGas Performance Specifications
SvO
2
Performance Specifications
SvO
2
Range 10% to 100%
Accuracy ± 2 % (i.e. ± 2 units), 1 standard deviation over 40 % to 100 %
range.
Resolution 1%
Stability (system) Drift < 2% over 24 hours
Response Time (10 % to 90%) 5 seconds
SvO
2
Alarm Specifications Range Adjustment Delay
SvO
2
10% to 100% 1% max. 15+4 seconds after value goes beyond the
low/high alarm limit settings