Philips MP40/50 Medical Alarms User Manual


 
4 Patient Alarms and INOPs Patient Alarm Messages
54
***BRADY (Pulse)
or
***BRADY xxx<yyy
Press, SpO
2
The heart rate from the Pulse signal has
fallen below the bradycardia limit. xxx
denotes the lowest measured value; yyy
is the bradycardia limit.
numeric flashes and alarm
limit is highlighted, red alarm
lamp, alarm tone.
**CCO/CCI HIGH CCO Continuous Cardiac Output or CC
Index is above the high alarm limit.
numeric flashes and high alarm
limit is highlighted, yellow
alarm lamp, alarm tone.
**CCO/CCI LOW CCO Continuous Cardiac Output or CC
Index is below the low alarm limit.
numeric flashes and low alarm
limit is highlighted, yellow
alarm lamp, alarm tone.
**CPP HIGH CPP The CPP value has exceeded the high
alarm limit.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone
**CPP LOW CPP The CPP value has fallen below the low
alarm limit.
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
***DESAT or
***DESAT xxx<yyy
SpO
2
The SpO
2
value has fallen below the
desaturation alarm limit. xxx denotes
the lowest measured value, and yyy is
the desaturation limit.
numeric flashes, red alarm
lamp, alarm tone.
**et <Agent label>
HIGH
AGM The end tidal agent high alarm limit has
been exceeded.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
**et <Agent label>
LOW
AGM The end tidal agent value has fallen
below the low alarm limit.
numeric flashes and low alarm
limit is highlighted, yellow
alarm lamp, alarm tone.
**etCO2 HIGH CO
2
, Resp,
AGM
The end tidal CO
2
high alarm limit has
been exceeded.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
**etCO2 LOW CO
2
, Resp,
AGM
The end tidal CO
2
value has fallen
below the low alarm limit.
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
**etO2 HIGH O
2
, AGM The end tidal O
2
high alarm limit has
been exceeded.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
**etO2 LOW O
2
, AGM The end tidal O
2
value has fallen below
the low alarm limit.
numeric flashes, and low limit
is highlighted, yellow alarm
lamp, alarm tone.
*/**/***
EVENT:<GRP>
Event
surveillance
An event has occurred and the event
notification is configured to alarm.
<GRP> is the event group
event group name flashes,
yellow or red alarm lamp and
alarm tone
*/**/*** EVENT
at Information center
Event
surveillance
An event has occurred and the event
notification is configured to alarm.
Check on the monitor for more details
on event group.
(on monitor) event group
name flashes, yellow or red
alarm lamp and alarm tone
***EXTREME BRADY ECG The bradycardia limit has been
exceeded.
numeric flashes and alarm
limit is highlighted, red alarm
lamp, alarm tone.
***EXTREME TACHY ECG The tachycardia limit has been
exceeded.
numeric flashes and alarm
limit is highlighted, red alarm
lamp, alarm tone.
Alarm Message From Condition Indication