Alarm Limits 3Alarms
47
Viewing All Alarm Limits
The Alarm Limits overview window lists the currently set alarm limits for all measurements. If an
Apnea alarm delay time is set, this is also shown. The Alarms Off symbol is shown beside the
measurement label of any measurement whose alarm switched off.
To open the Alarm Limits window, either select any alarm field to open the Alarm
Messages window, then select the Alarm Limits pop-up key, or select the Alarm
Limits SmartKey, if configured.
Select Show ST Limits to expand the list of ST leads and view the currently set alarm limits.
Selecting Hide ST Limits hides the list again.
You can use the pop-up keys that open with the Alarm Limits window to perform common tasks:
All Al. On/All Al. Off,
All Lim. Narrow/All Lim. Wide to set narrow or wide alarm AutoLimits for all
measurements.
Print Limits/Record Limits to print a list of all current alarm limit settings on a
connected printer or recorder.
Measurement labels, with alarms
off symbol where appropriate
Current alarm limits
Graphic view of current
yellow and red alarm
limits and currently
monitored measurement
value
Off indicates the
measurement is switched
off
Alarm Limits
ST-V4
ST-V5
ST-V6
SpO2
Pulse (SpO2)
ABPs
PAPd
awRR
NBPs
Apnea Time
Off