Philips MP40/50 Medical Alarms User Manual


 
10 Monitoring NBP Starting and Stopping Measurements
146
Meas Time: the time of the most recent NBP measurement, or
Next Meas: the time until the next measurement in an
automatic series, displayed with a graphic representation of the
remaining time, as shown here.
During measurements the cuff pressure is displayed instead of the units and the repeat time. An early
systolic value gives you a preliminary indication of the systolic blood pressure during measurement.
Starting and Stopping Measurements
Use the Setup menu, SmartKeys or the MMS hardkey to start and stop measurements.
CAUTION Use clinical judgement to decide whether to perform repeated series of STAT measurements because of
the risk of purpura, ischemia and neuropathy in the limb with the cuff.
Enabling Automatic Mode and Setting Repetition
Time
1 In the Setup NBP menu, select Auto/Man.
2 Toggle between Auto/Man, if necessary, to pick the measurement method.
3 If making an automatic measurement, select Repeat Time and set the time interval between two
measurements.
Action to be performed NBP Setup menu SmartKeys MMS hardkey
Start/Stop manual measurement
Start Auto series
Stop current automatic
measurement
Start/Stop
Start/Stop
Stop Automatic, Manual or STAT
measurement AND series
Stop All
---
Start STAT measurement
NBP STAT
STAT (for MMS
without Pressure/Temp
measurement)
Stop current STAT measurement
and end series
Start/Stop
Start/Stop
NBP STAT
STAT (for MMS
without Pressure/Temp
measurement)