Philips MP40/50 Medical Alarms User Manual


 
4 Patient Alarms and INOPs Patient Alarm Messages
56
*/**NON-SUSTAIN VT ECG/
Arrhythmia
A run of Vs having a ventricular HR>V-
Tach HR limit, but lasting for less than
the V-Tach Run limit has been detected.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
*/**PACER NOT CAPT ECG/
Arrhythmia
(paced
patients only)
A missed beat with a pace pulse was
detected.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
*/**PACER NT
PACING
ECG/
Arrhythmia
(paced
patients only)
A missed beat without a pace pulse was
detected.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
*/**PAIR PVCs ECG/
Arrhythmia
A non-ventricular contraction, followed
by two ventricular contractions,
followed by a non-ventricular
contraction has been detected.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
*/**PAUSE ECG/
Arrhythmia
No beat detected for a period greater
than the pause threshold.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
***<Pressure>
DISCONNECT
PRESS The pressure is non-pulsatile and the
mean pressure is continuously less than
10mmHg (1.3kPa). This alarm occurs
only with arterial pressures (P, ABP,
ART, Ao, BAP, FAP, PAP, UAP).
numeric flashes, red alarm
lamp, alarm tone.
**<Pressure> HIGH PRESS The measured pressure value is above
the high alarm limit. s, d, or m after the
label indicates whether the systolic,
diastolic or mean pressure has crossed
the limit.
numeric flashes, high limit is
highlighted, yellow alarm
lamp, alarm tone.
**<Pressure> LOW PRESS The measured pressure value is below
the low alarm limit. s, d, or m after the
label indicates whether the systolic,
diastolic or mean pressure has crossed
the limit.
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
**PULSE HIGH PRESS
SpO
2
The pulse rate has exceeded the high
alarm limit.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
**PULSE LOW PRESS
SpO
2
The pulse rate has dropped below the
low alarm limit.
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
*/**PVCs/min HIGH ECG/
Arrhythmia
More premature ventricular
contractions have been detected in a
minute than the limit.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
*/**R-ON-T PVCs ECG/
Arrhythmia
For HR <100, a PVC with R-R interval
< 1/3 the average interval followed by a
compensatory pause of 1.25 x average
R-R interval or two such Vs without
compensatory pause occurring within 5
minutes of each other. (When HR
>100, 1/3 R-R interval is too short for
detection.).
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
Alarm Message From Condition Indication