Philips MP40/50 Medical Alarms User Manual


 
24 Recording Reloading Paper
264
Reloading Paper
M1116B
1
Use the latch on the right side of the recorder door to pull the door open.
2 Remove the empty core.
3 Insert a new roll so that it fits snugly into its housing and the paper end is
feeding from the top.
4 Pull out some paper and fold along the front edge at a 45° angle. This makes it easier to feed the
paper under the roller as shown.
5 Feed the paper through and pull some paper out from the top of the roller.
6 Close the recorder door.
7 To test if paper is loaded correctly, start a recording. If no printing appears,
paper may be loaded backwards. Try reloading the paper.
MP20/MP30 1 Use the latch on the right side of the recorder door to pull the door open.
2 Remove the empty core.
3 Insert a new roll and secure it in place on the paper holder. The paper
feeds from the bottom of the roll and over the top of the recorder door.
4 With at least one inch of paper extending beyond the edge of the door,
swing the recorder door up and push it firmly closed.
5 To test if paper is loaded correctly, start a recording. If no printing appears, paper may be loaded
backwards. Try reloading the paper.
CAUTION When the recorder is disabled (by removal from its slot, door open, or out of paper), any alarm
recordings will be sent to the central station recorder, if there is one. If no recorder is available, alarm
recordings may be lost during the time the recorder is disabled. The message
no alarm recording
available
will be displayed.