Philips MP40/50 Medical Alarms User Manual


 
5 Managing Patients Care Groups
90
Telemetry Data Overview Screen Element
WARNING Waves and numeric information on the telemetry screen element are delayed for several seconds. If you
need realtime data, for example for defibrillation, always use the monitor ECG instead of telemetry.
Silencing Telemetry Alarms from the Bedside
When a telemetry device is paired with the monitor, an audible status change notification is issued for
new telemetry alarms. The volume can be configured.
If configured, alarms generated from a paired telemetry device can be silenced at the bedside.
To silence telemetry alarms from the bedside,
1 select the telemetry screen element
2 select the Silence Bed pop-up key
Depending on your configuration, the Silence Bed key may silence both telemetry alarms and
bedside alarms.
Suspending Telemetry Alarms
When you select Pause Alarms or Alarms Off at the monitor, the alarms are off or paused for
the bedside measurements.
Refer to the Information Center Instructions for Use for the Suspend/Pause alarms behavior of the
Information Center and telemetry device.
Using Standby
When you select Standby mode at the monitor, the bedside goes into Standby mode.
Refer to the Information Center Instructions for Use for the Standby behavior of the Information
Center and telemetry device.
Te l em e tr y
data overview
screen element
Bed1
Bed4 Bed5
Bed9
Room1
Bed7
Room3
Room1
Room4
TelemetryProfiles
23Feb 17:51Mayer, John AdultRoom3
Tele 3
NBP mmHg
HR
PVC
MCL
1mV
1mV
Delayed
Pleth
CO
2
Pulse
Perf
etCO
2
mmHg
Temp
awRR rpm