Philips MP40/50 Medical Alarms User Manual


 
34 Installation and Specifications Measurement Specifications
354
Invasive Pressure and Pulse
Complies with IEC 60601-2-34:2000/EN60601-2-34:2000.
NBP Overpressure Settings
Adult > 300 mmHg (40 kPa) > 2 sec not user adjustable
Pedi > 300 mmHg (40 kPa) > 2 sec
Neo > 150 mmHg (20 kPa) > 2 sec
Invasive Pressure Performance Specifications
Measurement Range –40 to 360 mmHg
Pulse Rate Range 25 to 350 bpm
Accuracy
±1% Full Range
Resolution
1 bpm
Input Sensitivity Sensitivity:5µV/V/mmHg (37.5µV/V/kPa)
Adjustment range:±10%
Transducer Load Impedance:200 to 2000 (resistive)
Output Impedance:3000 (resistive)
Frequency Response dc to 12.5 Hz or 40 Hz
Zero Adjustment Range: ±200 mmHg (±26 kPa)
Accuracy
±1 mmHg (±0.1 kPa)
Drift
Less than 0.1mmHg/°C (0.013 kPa/°C)
Gain Accuracy Accuracy ±1%
Drift
Less than 0.05%/°C
Non linearity
and Hysteresis
Error of 0.4% FS (@CAL 200 mmHg)
Overall Accuracy (including
transducer)
± 4% of reading or ± 4 mmHg (± 0.5 kPa), whichever is greater
Analog Output
available only with
M1006B #C01 (@
CAL 200 mmHg)
Range -0.4 V to 3.6 V
Level
1 V / 100 mmHg
Accuracy
± 3% full scale
Offset
± 30 mV
Resolution
8 Bit (@ 5 V range)
Signal delay
20 ms
Volume displacement of CPJ840J6 0.1 mm
3
/100 mmHg