Philips MP40/50 Medical Alarms User Manual


 
34 Installation and Specifications Monitor Performance Specifications
344
Events Information trigger condition and time, event classification and associated
detailed view of episode data
Episode data
configurable (except MP20/30), either:
4 minutes of high resolution trend or
20 minutes of numerics trend @ 12 sec. resolution or
15 seconds of 4 waves @ 125 samples/sec. (Snapshot)
including all current numerics, alarms and inops
For MP20/30: 4 minutes of high resolution trend
Capacity (max)
25 or 50 events for either 8 or 24 hours
Alarm signal System delay less than 3 seconds
Pause duration
1,2,3 minutes or infinite, depending on configuration
Extended alarm pause
5 or 10 minutes
Review Alarms Information all alarms / inops, main alarms on/off, alarm silence and time
of occurrence
Capacity
100 items
Real Time Clock Range from: January 1, 1997, 00:00 to: December 31, 2080, 23:59
Accuracy
< 2 seconds per day (typically)
Hold Time
infinite if powered by AC; otherwise at least 48 hours (typical:
> 72 hours)
Buffered Memory Hold Time if powered by AC: infinite
without power: at least 48 hours (typical: > 72 hours)
Contents
Active settings, trends, patient data, realtime reports, events,
review alarms
MMS M3001A Performance Specifications
Tre n d s Trend Data Buffered
Memory
if powered by monitor mains connection via MSL: infinite
without power applied: at least 6 hours
Contents
trend data, patient identification and all active settings
D80 Intelligent Display Performance Specifications
Power Specifications Power consumption < 145 W
Line Voltage 100 to 240 V ~
Current 1.6 to 0.7 A
Frequency 50/60 Hz ~
Docking Station M8045A Performance Specifications
Power Line Voltage 100 - 240V
Frequency
50/60 Hz
Power Consumption
negligible (for Power On LED)
Current to attached
patient monitor
max. 3.5 A
MP20/30, MP40/50, MP60/70/80/90