Philips MP40/50 Medical Alarms User Manual


 
Pulse Default Settings 35 Default Settings Appendix
373
Pulse Default Settings
Pulse alarms use the settings of the currently selected Pulse alarm source.
Lead I, II, III, V, aVR, aVL, aVF, V1-6,
MCL Settings
Factory Adult Factory Pedi Factory Neo
ST
(Label) On Off
For Alarm Mode = Single-ST
ST
(Label) High +2.0 mm
ST
(Label) Low -2.0 mm
For Alarm Mode = Multi-ST
ST
(Label) High +1.0 mm
ST
(Label) Low -1.0 mm
Pulse Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Alarm Source HR Auto
Pulse (Label) On
System Pulse SpO2 Auto
Alarms Off Enabled
Alarm Source Sel. Enabled
Pulse Alarm Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Pulse (SpO2) on
Pulse Alarms on
High Limit 120 bpm 160 bpm 200 bpm
Low Limit 50 bpm 75 bpm 100 bpm
Extr Brady 20 bpm
Brady Clamp 40 bpm 40 bpm 50 bpm
Extr Tachy 20 bpm
Tachy Clamp 200 bpm 220 bpm 240 bpm