Philips MP40/50 Medical Alarms User Manual


 
35 Default Settings Appendix CO
2
Default Settings
378
CO
2
Default Settings
Spirometry Default Settings
CO
2
Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
etCO
2
low 30 25
etCO
2
high 50 60
imCO
2
high 4
CO
2
Alarms on
Unit mmHg
Scale 40 mmHg 50
ImCO
2
on Off
N
2
O Corr (only
M3015A/
M3016A)
Off
Oxygen Corr (only
M3014A)
16%
Gas Corr (only
M3014A)
Off
Humidity Corr BTPS
Max Hold Off
AwRR On
AwRR Alarms On
AwRR high limit 30 100
AwRR low limit 8 30
Apnea time 20 secs
Color Yellow White
Spirometry
Settings
Adult Pedi Neo
APNEA high limit 20 20 20
MVexp low limit 4.0 2.5 0.4
MVexp high limit 8.0 4.0 0.8
PIP high limit 40 25 20
PEEP low limit 0 0 0
PEEP high limit252525
RRspir low limit 8 8 30
RRSpir high limit303060
AWF scale
±150 l/min ±100 l/min ±20 l/min
AWV scale
±800 ml ±200 ml ±50 ml
AWP scale
±40 cmH
2
O ±40 cmH
2
O ±20 cmH
2
O