Philips MP40/50 Medical Alarms User Manual


 
NBP Default Settings 35 Default Settings Appendix
375
NBP Default Settings
Temperature Default Settings
Pulse Alarms On/Off on on on
Pulse High Limit 120 bpm 160bpm 200 bpm
Pulse Low Limit 50 bpm 75 bpm 100 bpm
Extr Brady 20 bpm 20 bpm 20 bpm
Brady Clamp 40 bpm 40 bpm 50 bpm
ExtrTachy 20 bpm 20 bpm 20 bpm
Tachy Clamp 200 bpm 220 bpm 240 bpm
SpO
2
Alarm Default Settings
Setting adult pediatric neonatal
NBP Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Auto/Manual Auto Manual
Alarms from Sys.
High Alarm Limit 160/ 90 (110) 120/ 70 (90) 90/ 60 (70) 180/ 90 (110)
Low Alarm Limit 90/ 50 (60) 70/ 40 (50) 40/ 20 (24) 70/ 50 (65)
Alarms On
NBP On
Repetition Time 15 min 3 min
Unit mmHg
Done Tone Off On
Start Time Synchronized NotSynchron.
VP Pressure 60 mmHg 40 mmHg 30 mmHg
Reference Auscultatory Invasive
Color Red Magenta
Temp Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Low Limit 36 35
High Limit 39
Alarms On
Unit
0
C
Range 35...43 30 ... 43
Color Green Light Green