Philips MP40/50 Medical Alarms User Manual


 
24 Recording Recording Priorities
262
Recording Priorities
Manually-started recordings have priority over automatically-started recordings. If an
automatically-triggered alarm recording is running, and a realtime or delayed recording is manually
started, the alarm recording is stopped and the manually-requested recording is started.
More recent manually-started recordings have priority over older manually-started recordings. If a
manually-started recording is running, and another manually-started recording is triggered, then the
older recording is stopped and the more recent manually-started recording is started.
Alarm recordings are prioritized according to alarm priority. If an alarm recording triggered by a
yellow alarm is running and a new alarm recording is triggered by a red alarm, the yellow alarm
recording is stopped and the red alarm recording is started.
Sample Recording Strip
The information printed on the recording strip includes the patient name and MRN, bed number,
date and time of recording, recording speed, and recording code. Active alarm and INOP messages as
well as numerics for all currently monitored measurements are also printed.
Recording strip annotations are printed at the beginning of the recording strip and updated at regular
intervals, every 15 minutes for recordings made at speeds lower than 6.25 mm/s, and every 60 seconds
for recordings made at speeds greater or equal to 6.25 mm/s.
This sample recording strip shows a typical initial annotation:
Recording Strip Code
The recording strip code printed in the first line of the initial annotation has up to seven characters,
specifying recording type, operating mode, application area, patient category, and delay time, if
applicable.
90DIN 25 mm/sec
23 Jan 02 10:43 Bed 3
Green, Sarah 37549201
*** EXTREME BRADY
** ABPs HIGH
** PAPd HIGH
CCO NO CALIBRATION
SOME ECG ALARMS OFF
ABP REDUCE SIZE
HR 120 bpm
SPO2 95%
Pulse 120 bpm
PERF 10.0
ABP 120/70 (91) mmHg
PAP 28/15 /21) mmHg
NBP 120/80 (90) mmHg 23 Jan 02
10:31
C.O.
C.I.
CCO -?- l/min
CCI
etCO2 40 mmHg
imCO2 0 mmHg
awRR
RR 15 rpm
Tnaso 37.0×C
Tblood 37.0×C
SV 94 ml
SI 48 ml/m2
Recording
strip code
Patient
information
and medical
record no.
Alarms
and
INOPs
Recording speed
Measurement numerics