Philips MP40/50 Medical Alarms User Manual


 
M4605A Battery Specifications 34 Installation and Specifications
341
M4605A Battery Specifications
MP20/MP30/
MP40/MP50
Only
Two batteries are required to operate the MP40/MP50 monitors. The MP20/MP30 monitors can
operate with one battery.
Humidity Range Operating 95% Relative Humidity (RH) max @ 40°C (104°F) (non
condensing)
Non-operating
95% Relative Humidity (RH) max @ 65°C (150°F) (non
condensing)
Altitude Range Operating up to 3048 m (10000 ft)
Non-operating
up to 3048 m (10000 ft)
Remote SpeedPoint M8026A
Item Condition Range
Temperature range Operating 0 ... 55 °C (32 ...130 °F)
Storage
-20 … 60 °C (-4 ... 140 °F)
Humidity range Operating 95 %RH max. at 40 °C (100 °F)
Storage
85 %RH max. at 50 °C (120 °F)
Altitude range Operating -500 … 4.600 m (-1600 ... 15000 ft)
Storage
-500 … 13.100 m (-1600 ... 43000 ft)
Remote Alarm Device M8025A
Item Condition Range
Temperature range Operating 0 ... 55 °C (32 ...130 °F)
Storage
-20 … 60 °C (-4 ... 140 °F)
Humidity range Operating 95 %RH max. at 40 °C (100 °F)
Storage
85 %RH max. at 50 °C (120 °F)
Altitude range Operating -500 … 4.600 m (-1600 ... 15000 ft)
Storage
-500 … 13.100 m (-1600 ... 43000 ft)
Thermal Array Recorder Module M1116B
Item Condition Range
M4605A Battery Specifications
Physical Specifications
W x D x H 149 mm (5.866 in) x89 mm (3.504 in) x 19.8 mm (0.78 in)
Weight 490 g (1.08 lb) per battery
Performance Specifications
Nominal Voltage 10.8 Volt
Rated Capacity at discharge C/5 6000 mAh
Continuous Discharge Capability 6.5 A
Environmental Specifications