Philips MP40/50 Medical Alarms User Manual


 
35 Default Settings Appendix EEG Default Settings
380
EEG Default Settings
BIS Default Settings
EEG Settings Factory Adult Factory Pedi Factory Neo
TP, SEF On
MDF, PPF, Delta, Theta,
Alpha, Beta
Off
SEF Threshold 90 %
Numeric Average 8 sec
Wave Scale 100uV (or +/- 50uV if Show Gridlines is configured to Yes)
Show Gridlines No
Low Filter 0.5 Hz
High Filter 30 Hz
Impedance Limit 5 kOhm
Smoothing CSA On
Color Yellow
BIS Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
SQI On
EMG On
SR On
SEF Off
TP Off
Bursts On Off
Scale 100uV (or +/- 50uV
if Show Gridlines
is configured to Yes)
Show Gridlines No
Filters On
Low Filter 2 Hz
High Filter 70 Hz
Notch Filter On
High Alarm Limit 70
Low Alarm Limit 20
Alarms On
Smoothing Rate 30 sec 15 sec
Color Yellow Magenta