SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING
Part No 1114842 21 2G Tarsys®Powered Seating System
Post-Service Inspection Checklist
ƽ WARNING
NEVER operate the wheelchair while in any tilted/reclined/back angle position over
20° relative to the vertical position. If the drive lock-out does not stop the
wheelchair from operating in a tilt position over 20° relative to the vertical position,
DO NOT operate the wheelchair. The drive lock-out settings MUST be adjusted.
Refer to Reference Documents on page 2.
Cables MUST be secured to the wheelchair frame and/or base with tie-wraps after
servicing is complete. Failure to follow the warnings and instructions below could
result in injury to the users, attendants and/or bystanders and/or damage to the
wheelchair.
Cables MUST be secured so there are no loops of excess cable extending away from
the wheelchair. Bundle all excess cable together and secure with a tie-wrap. It may
also be necessary to secure these bundles to the frame and/or base.
ALWAYS test all wheelchair functions after securing the cables to be sure cables
DO NOT get pinched or crushed during operation of the wheelchair.
The following checklist must be performed after any procedure that requires:
Removing/installing the seat
Removing/installing/adjusting the tilt links
Replacing electronic components
Replacing actuators
Verify all attaching hardware is tightened securely.
Make sure drive lock-out engages when the wheelchair is in any tilt/recline/back angle
combination over 20° relative to the vertical position.
Cycle system up and down to verify wiring harnesses DO NOT obstruct the path of
the system. If they do, perform one of the following:
Wires were damaged during inspection - Replace damaged wires.
Wires were not damaged during inspection - Cut tie-wraps and relocate wires to a
location where they will not become damaged.
Tilt Systems - Verify tilt link is properly secured to the frame.
Tilt Systems - Ensure there is no interference with the base frame when the system is
fully tilted.
With the wheelchair unoccupied, test all system functions to verify proper operation.
Verify charger function.
Install all shrouds.
Make sure linear guides DO NOT bind.