Invacare 2GT Wheelchair User Manual


 
SECTION 13—TILT AND RECLINE
2G Tarsys®Powered Seating System 134 Part No 1114842
FIGURE 13.6 Removing/Installing the Recline Actuator
Testing the Actuators
1. Cycle each function of the seating system (tilt, recline, VSR, power legs) through the
full range of motion and observe the cables to make sure no cables are pinched,
crushed or pulled by the mechanisms.
2. Flip back armrests and observe the cables to make sure the joystick and/or switch
cables are free to move and not pulled.
3. Perform one of the following:
If damage to the cables is found, replace the cables and re-route the new cables so
they DO NOT get damaged during wheelchair/seating system operation.
If the cables are in danger of becoming damaged, re-route the cables to ensure
damage does not occur.
4. Secure the cables. Refer to Securing the Cables on page 157.
5. Perform the post-service inspection. Refer to Post-Service Inspection Checklist on
page 21.
Recline
Actuator
Locknut
(1 of 4)
Screw
Actuator
Bracket
(1 of 2)
Socket Head
Screw
(1 of 4)
Locknut
Pin
Seat Frame Assembly
NOTE: The crossbeam is
shown removed for clarity.
Crossbeam