Invacare 2GT Wheelchair User Manual


 
SECTION 13—TILT AND RECLINE
Part No 1114842 133 2G Tarsys®Powered Seating System
Installing
1. Position the new VSR actuator on the back as shown in FIGURE 13.5.
2. Line up the mounting holes in the new VSR actuator and VSR actuator bracket.
3. Install the hex bolt, washers and locknut that secure the new VSR actuator to the VSR
actuator mounting bracket. Tighten securely. Torque to 75 ft-lbs ± 20%. Refer to
FIGURE 13.5 for hardware orientation.
4. Connect the VSR actuator connector to the TRCM or TAC VSR actuator connectors
(Not shown).
5. Reinstall the VSR actuator shroud. Refer to Removing/Installing VSR Shroud
on
page 100.
6. Perform the post-service inspection. Refer to Post-Service Inspection Checklist
on
page 21.
Removing/Installing the Recline Actuator
NOTE: For this procedure, refer to FIGURE 13.6 on page 134.
Removing
1. Remove the locknut and screw that secure the front of the recline actuator to the seat
frame assembly.
2. Remove the four socket head screws and locknuts that secure the right and left
actuator brackets to the crossbeam.
3. Take note of the right and left actuator bracket orientations.
NOTE: The left and right brackets have long and short sides. The right and left brackets must be
put back correctly or range of motion will be lost.
4. Release the pins that secure the actuator brackets to the recline actuator.
5. Remove the recline actuator.
Installing
1. Secure the pins in the right and left actuator brackets to the recline actuator. Make sure
the orientation of each bracket is correct.
2. Install the right and left actuator brackets, with the recline actuator attached, to the
crossbeam assembly with the four socket head screws and locknuts.
3. Install the front of the recline actuator to the seat frame assembly with the locknut and
screw.