Invacare 2GT Wheelchair User Manual


 
SECTION 9—POWER ELEVATING LEGRESTS
Part No 1114842 107 2G Tarsys®Powered Seating System
Adjusting the Footplate Depth
NOTE: For this procedure, refer to
FIGURE 9.3.
1. Remove the two flat head screws and
the two barrel nuts that secure the
footplate to the footplate clamp.
2. Reposition the footplate on the
footplate clamp.
3. Align the depth adjustment holes on
the footplate to the mounting holes on
the footplate clamp.
FIGURE 9.3 Adjusting the Footplate Depth
4. Insert the barrel nuts through the bottom of the footplate clamp.
5. Install the two flat head screws through the footplate and footplate clamp.
6. Securely tighten with two barrel nuts.
7. Repeat STEPS 1-6 for opposite footplate if necessary.
Adjusting the Footplate Angle
NOTE: For this procedure, refer to FIGURE 9.4 on page 108.
1. Loosen, but DO NOT remove, the two flat head screws and the two barrel nuts that
secure the footplate to the footplate clamp. Refer to Detail “A”.
2. Rotate the footplate clamp on the pivot hinge until the desired angle is achieved. Refer
to Detail “B”.
3. Securely tighten footplate to footplate clamp and hinge pivot with the two flat head
screws and barrel nuts.
4. Repeat STEPS 1-3 for opposite side if necessary.
5. For additional angle adjustment, perform the following:
A. Loosen the set screw.
B. Rotate footplate up or down to desired position.
C. Retighten set screw.
Footplate Clamp
Flat Head Screws
Depth
Adjustment
Holes
Barrel Nuts
Footplate
Mounting Holes