Hitachi S-4800 Microscope & Magnifier User Manual


 
3-80
3.8 At the End of Daily Use
3.8.1 Turning High Voltage Off
(1) Close opening dialog windows. Save data if necessary
(2) Click the HV OFF button
on the control panel.
3.8.2 Setting the Stage at the Specimen Exchange Position
Set the all 5 axes of the specimen stage at the exchange position.
(1) Type1 manual stage
Release the stage lock.
Set X = 25.0, Y = 25.0, Z = 8.0 Tilt = 0 and R = 0.
(2) Type2 motorized stage
Release the stage lock.
Click EXC button on the control panel. Wait until the indicator of the button changed to green
color.
3.8.3. Taking Out a Specimen
(1) Press the OPEN button on the specimen exchange chamber. Once the exchange chamber
will be evacuated and then, the gate valve will open.
(2) Turn the knob of the specimen exchange rod clockwise so as the Unlock mark comes
upwards.
Insert the rod straightly and plug the two spring pins at the end of the rod to holes of the
specimen holder.