4535 612 34161 HD3 Service Manual Page 98
CSIP: Level 2 Installation: Physical Installation Procedures
Figure 5-10 Rolling the System Off the Pallet
10. Ensure the ramp is properly aligned
in front of the crate.
11. Unlock the front casters.
12. Lift the antistatic bag and position
yourself on one side of the crate
toward the front of the ultrasound
system.
13. Grasp the control panel handle and
carefully guide the system down the
ramp and onto the floor.