Philips 612 Ultrasonic Jewelry Cleaner User Manual


 
4535 612 34161 HD3 Service Manual Page 78
CSIP Level 1 Installation: Powering On and Configuring the System
Configuring the
DICOM Feature
If the system you are installing has the DICOM option, you or the customer will need to config-
ure the system DICOM feature. Since configuring the system for DICOM operation requires
specific information from the customer network administrator (see “Gathering Site Network
(DICOM) Information” on page 69), Philips recommends that you work with the customer to
configure the DICOM feature with the specific values that need to be entered. At initial installa-
tion, refer the customer to the HD3 Getting Started manual, which has instructions for viewing,
setting, or changing DICOM information and gives a description of each field in the "Glossary."
The basic steps for initially configuring the system for DICOM communications over a network
are as follows:
1. Enter the system DICOM parameters:
- Enter the AE Title, Station Name, and Port number for the HD3 system (page 78).
- Enter the network settings for the HD3 system (page 80).
2. Identify the DICOM storage devices (servers) that the system will use (page 80).
3. Identify the DICOM printers that the system will use (page 80).
4. After the DICOM feature has been configured, you may want to test the communication
between the system and a server or a printer that was entered.(page 83).
5. Configure the DICOM Log file settings (page 83).
To configure the system DICOM parameters
1. Press Setup.
The Setup window opens (Figure 5-1).
2. Click the DICOM tab.
The DICOM setup window opens (Figure 5-3).