CSIP Level 1 Installation: Inspecting the Installation Site
Install the system power cord (page 72). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Start up and configure the system (page 72). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Set or verify the institution name, user language, and time and date settings (page 73 through
page 75). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Verify system options (page 76). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configure the DICOM/connectivity network, if required (page 77).. . . . . . . . . . . . . . . . . .
Configure other system settings (optional) (page 87). . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Prepare the peripherals (page 87). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Perform a system functional check (page 87). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Present the system (page 88). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Inspecting the
Installation
Site
Conduct an inspection of the site prior to system installation to ensure a trouble-free installation
and to assist the customer in providing an environment that supports reliable performance of the
system.
Checking
Physical Access
for Delivery and
Installation
Check that the installation site is physically acceptable as follows:
With the customer, evaluate site access for delivery. Consider the availability of a loading
dock, elevators, ramps, and the width of passageways and doors.
Verify that there is adequate space available for system uncrating and installation.
At each location where the system will be operated, verify that there is enough space for the
system and any external peripherals.
Check adjacent rooms for activity and types of equipment that might negatively affect the
ultrasound system.