CSIP Level 1 Troubleshooting: Viewing System Information and Logs
To display system information
1. Press Set Up (F10).
2. Click the Admin tab.
3. In the Service area, click Show to display system information.
4. Use the scroll bars to view the system information.
5. Click Close when done.
To display the keystroke or DICOM log
NOTE You can view only one keystroke log at a time. However, you can copy or delete multi-
ple keystroke logs.
1. Press Set Up (F10).
2. Click the Admin tab.
3. In the Service area, click Open to open the keystroke or DICOM log.
4. In the dialog box, do the following:
- Enter the number of days a log file should be active. For example, if you enter 31 days,
the log file is deleted when it becomes older than 31 days.
- Enter the maximum size of the log file in kilobytes.
5. To view a log file, select the log file and click View selected file.
6. To copy a log file or files, select the file or files, select the removable media destination, and
click Copy selected files.
7. To delete a file after copying, select the Delete files after copy check box.