CSIP Level 1 Installation: Powering On and Configuring the System
To test the communication between the system and a server or a printer
1. Press Setup.
The Setup window opens (Figure 5-1).
2. Click the DICOM tab.
The DICOM setup window opens (Figure 5-3).
3. Highlight the device that you want to test.
4. Click Test to initiate a DICOM verify (“ping”) test signal from the system to the device.
The test will produce a results message similar to one of the examples shown in Figure 5-6.
To configure the DICOM Log file
NOTE The defaults for the DICOM logging parameters are reasonable for most circum-
stances. Do not change the settings unless you have a specific troubleshooting need.
1. Press Setup.
The Setup window opens (Figure 5-1).
2. Click the Admin tab.
The Admin setup window opens (Figure 5-7).
3. Click (DICOM Log) Open.
The DICOM Log window opens (Figure 5-8).
4. Enter the DICOM Log parameters and click Close.
5. Click Save on the the Admin tab.