4535 612 34161 HD3 Service Manual Page 204
CSIP Level 2 Disassembly: Disassembly (Removal) Procedures
System Casters
Figure 10-50 Preparing the System for Caster
Replacement
3
2
To remove a system casters
1. Disconnect the system power cord
from the main AC supply.
WARNINGThe following steps
require that the system
be tipped over to the
floor. To reduce the risk
of injury or system dam-
age from toppling, tilt the
monitor down onto the
control panel surface and
lower the control panel
to its lowest position.
2. Lower the control panel to its lowest
position.
3. Tilt the monitor down onto the con-
trol panel.