Transmitter default values and ranges (continued)Table D-1:
Type Parameter Default Range Comments
Base volume time sec
Volume flow conversion factor 1
Variable map-
ping
Primary variable Mass flow
Secondary variable Density
Tertiary variable Mass flow
Quaternary variable Volume flow
mA output 1 Primary variable Mass flow
LRV –200.00000 g/
s
URV 200.00000 g/s
AO cutoff 0.00000 g/s
AO added damping 0.00000 sec
LSL –200 g/s Read-only.
LSL is calculated based on the
sensor size and characterization
parameters.
USL 200 g/s Read only.
USL is calculated based on the
sensor size and characterization
parameters.
MinSpan 0.3 g/s Read-only.
Fault action Downscale
AO fault level – downscale 2.0 mA 1.0 – 3.6 mA
AO fault level – upscale 22 mA 21.0 – 24.0 mA
Last measured value timeout 0.00 sec
mA output 2 Secondary variable Density
LRV 0.00 g/cm3
URV 10.00 g/cm3
AO cutoff Not-A-Number
AO added damping 0.00000 sec
LSL 0.00 g/cm
3
Read-only.
LSL is calculated based on the
sensor size and characterization
parameters.
USL 10.00 g/cm
3
Read only.
USL is calculated based on the
sensor size and characterization
parameters.
MinSpan 0.05 g/cm
3
Read-only.
Default values and ranges
Configuration and Use Manual 237