For curved-tube sensors, set Flow Cal (Flow Calibration Factor).
3. Set the density characterization parameters.
For straight-tube sensors, set D1, D2, DT, DTG, K1, K2, FD, DFQ1, and DFQ2.
For curved-tube sensors, set D1, D2, TC, K1, K2, and FD. (TC is sometimes shown
as DT.)
2.4.1 Sources and formats for characterization parameters
Different sensor tags display characterization parameters differently. On older sensors, the
sensor tag may not contain all the required parameters.
Sample sensor tags
Sample sensor tags are shown in the following illustrations:
Older curved-tube sensors (all sensors except T-Series): see Figure 2‐1
Newer curved-tube sensors (all sensors except T-Series): see Figure 2‐2
Older straight-tube sensors (T-Series): see Figure 2‐3
Newer straight-tube sensors (T-Series): see Figure 2‐4
Tag on older curved-tube sensors (all sensors except T-Series)Figure 2-1:
Quick start
10 Micro Motion
®
Model 1500 Transmitters with Analog Outputs