Emerson 1500 Pacemaker User Manual


 
7 Completing the configuration
Topics covered in this chapter:
Test or tune the system using sensor simulation
Back up transmitter configuration
Enable write‐protection on the transmitter configuration
7.1 Test or tune the system using sensor
simulation
Use sensor simulation to test the system's response to a variety of process conditions,
including boundary conditions, problem conditions, or alarm conditions, or to tune the
loop.
Restriction
Sensor simulation is available only on flowmeters with the enhanced core processor.
Prerequisites
Before enabling sensor simulation, ensure that your process can tolerate the effects of the
simulated process values.
Procedure
1. Navigate to the sensor simulation menu.
Communications tool Menu path
ProLink II
ProLink > Configuration > Sensor Simulation
ProLink III
Device Tools > Diagnostics > Testing > Sensor Simulation
Field Communicator
Service Tools > Simulate > Simulate Sensor
2. Enable sensor simulation.
3.
For mass flow, set Wave Form as desired and enter the required values.
Option Required values
Fixed
Fixed Value
Sawtooth
Period
Minimum
Maximum
Completing the configuration
Configuration and Use Manual 101