Emerson 1500 Pacemaker User Manual


 
Appendix D
Default values and ranges
D.1 Default values and ranges
The default values and ranges represent the typical factory transmitter configuration.
Depending on how the transmitter was ordered, certain values may have been configured
at the factory and are not represented in the default values and ranges.
Transmitter default values and rangesTable D-1:
Type Parameter Default Range Comments
Flow Flow direction Forward
Flow damping 0.8 sec
(1)
0.0 – 60.0 sec User-entered value is corrected
to nearest lower value in list of
preset values. In Special mode,
the preset values are 1/5 nor-
mal. For gas applications,
Micro Motion recommends a
minimum value of 2.56.
Flow calibration factor 1.00005.13 For T-Series sensors, this value
represents the FCF and FT fac-
tors concatenated.
Mass flow units g/s
Mass flow cutoff 0.0 g/s Recommended setting is 5% of
the sensor’s rated maximum
flowrate.
Volume flow type Liquid
Volume flow units L/s
Volume flow cutoff 0/0 L/s 0.0 – x L/s x is obtained by multiplying the
flow calibration factor by 0.2, us-
ing units of L/s.
Meter factors Mass factor 1
Density factor 1
Volume factor 1
Density Density damping 1.6 sec 0.0 – 60.0 sec User-entered value is corrected
to nearest value in list of preset
values.
(1) In Special mode, the default value is 0.64 sec.
Default values and ranges
Configuration and Use Manual 235