• If LD Optimization is enabled on your meter, disable it. To do this, choose Device Tools >
Configuration > LD Optimization. LD Optimization is used only with large sensors in
hydrocarbon applications. In some installations, only Micro Motion customer service
has access to this parameter. If this is the case, contact Micro Motion before
continuing.
• The calibrations must be performed without interruption, in the order shown. Make
sure that you are prepared to complete the process without interruption.
• Before performing the calibration, record your current calibration parameters. You
can do this by saving the current configuration to a file on the PC. If the calibration
fails, restore the known values.
Restriction
For T-Series sensors, the D1 calibration must be performed on air and the D2 calibration must be
performed on water.
Procedure
See Figure 9‐2.
D1 and D2 density calibration using ProLink IIIFigure 9-2:
Enter density of D1 fluid
D1 calibration
Close shutoff valve
downstream from sensor
Fill sensor with D1 fluid
Done
Device Tools >
Calibration >
Density Calibration >
Density Calibration – Point 1 (Air)
Close
Start Calibration
Enter density of D2 fluid
D2 calibration
Fill sensor with D2 fluid
Device Tools >
Calibration >
Density Calibration >
Density Calibration – Point 2 (Water)
Close
Start Calibration
Postrequisites
If you disabled LD Optimization before the calibration procedure, re-enable it.
Measurement support
132 Micro Motion
®
Model 1500 Transmitters with Analog Outputs