Emerson 1500 Pacemaker User Manual


 
10.3 Flow measurement problems
Flow measurement problems and recommended actionsTable 10-3:
Problem Possible causes Recommended actions
Flow indication at no
flow conditions or
zero offset
Misaligned piping (especially in new in-
stallations)
Open or leaking valve
Incorrect sensor zero
Verify that all of the characterization parame-
ters match the data on the sensor tag.
If the flow reading is not excessively high, re-
view the live zero. You may need to restore
the factory zero.
Check for open or leaking valves or seals.
Check for mounting stress on the sensor
(e.g., sensor being used to support piping,
misaligned piping).
Contact Micro Motion.
Erratic non-zero flow
rate under no-flow
conditions
Leaking valve or seal
Slug flow
Plugged or coated flow tube
Incorrect sensor orientation
Wiring problem
Vibration in pipeline at rate close to sen-
sor tube frequency
Damping value too low
Mounting stress on sensor
Verify that the sensor orientation is appropri-
ate for your application (refer to the sensor
installation manual).
Check the drive gain and the pickoff voltage.
See Section 10.26 and Section 10.27.
If the wiring between the sensor and the
transmitter includes a 9-wire segment, verify
that the 9-wire cable shields are correctly
grounded.
Check the wiring between the sensor and
transmitter. See Section 10.10.
For sensors with a junction box, check for
moisture in the junction box.
Purge the flow tubes.
Check for open or leaking valves or seals.
Check for sources of vibration.
Verify damping configuration.
Verify that the measurement units are con-
figured correctly for your application.
Check for slug flow. See Section 10.25.
Check for radio frequency interference. See
Section 10.13.
Contact Micro Motion.
Troubleshooting
156 Micro Motion
®
Model 1500 Transmitters with Analog Outputs