Important
Consult Micro Motion before performing a temperature calibration. Under normal circumstances,
the temperature circuit is stable and should not need an adjustment.
Procedure
See Figure 9‐7 and Figure 9‐8.
Temperature calibration using ProLink IIFigure 9-7:
Enter temperature of low-
temperature fluid
Temperature Offset calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with low-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Enter temperature of high-
temperature fluid
Temperature Slope calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with high-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Done
ProLink Menu >
Calibration >
Temp offset cal
ProLink Menu >
Calibration >
Temp slope cal
Measurement support
140 Micro Motion
®
Model 1500 Transmitters with Analog Outputs