Emerson 2200S Pacemaker User Manual


 
Configuration and Use Manual 57
Configuring Operational Parameters
Operation AppendicesMaintenance and TroubleshootingCommissioning
A013 Process too Noisy to Perform Auto Zero Fault Yes No
Zero Too Noisy
A014 Transmitter Failed Fault No No
Transmitter Failed
A016 Line RTD Temperature Out-Of-Range Fault Yes Yes
Line RTD Temperature Out-of-Range
A017 Meter RTD Temperature Out-Of-Range Fault Yes Yes
Meter RTD Temperature Out-of-Range
A020 Calibration Factors Unentered Fault Yes No
Calibration Factors Unentered (FlowCal)
A021 Unrecognized/Unentered Sensor Type Fault No No
Incorrect Sensor Type (K1)
A029 Internal Communication Failure Fault No No
PIC/Daughterboard Communication Failure
A030 Hardware/Software Incompatible Fault No No
Incorrect Board Type
A031 Undefined Fault No No
Low Power
A033 Tube Not Full Fault No Yes
Tube Not Full
A100 Primary mA Output Saturated Info Yes
(1)
No
Primary mA Output Saturated
A101 Primary mA Output Fixed Info Yes
(1)
No
Primary mA Output Fixed
A102 Drive Over-Range Info Yes No
Drive Overrange
A104 Calibration-In-Progress Info Yes
(1)
No
Calibration in Progress
A105 Slug Flow Info Yes No
Slug Flow
A106 Burst Mode Enabled Info Yes
(1)
No
Burst Mode Enabled
A107 Power Reset Occurred Ignore Yes No
Power Reset Occurred
A132 Simulation Mode Active Info Yes
(1)
No
Simulation Mode Active
(1) Can be set to either Info or Ignore, but cannot be set to Fault.
Table 5-5 Status alarms and severity levels continued
Alarm code
Communicator message
Default
severity Configurable
Affected by
LMV TimeoutProLink II message