Emerson 2200S Pacemaker User Manual


 
Configuration and Use Manual 125
Default Values and Ranges
Operation AppendicesMaintenance and TroubleshootingCommissioning Operation AppendicesMaintenance and TroubleshootingCommissioning Operation AppendicesMaintenance and TroubleshootingCommissioning Operation AppendicesMaintenance and TroubleshootingCommissioning
mA output Primary variable Mass flow
LRV –200.00000 g/s Automatically reset when
primary variable is reset.
URV 200.00000 g/s Automatically reset when
primary variable is reset.
AO cutoff 0.00000 g/s
AO added damping 0.00000 sec 0 – 440 seconds
Fault action Downscale
AO fault level – downscale 2.0 mA 1.0 – 3.6 mA
AO fault level – upscale 22 mA 21.0 – 24.0 mA
Last measured value timeout 0.00 sec
Display Update period 200 milliseconds 100 – 10,000
milliseconds
Variable 1 Mass flow rate
Variable 2 Mass total
Variable 3 Volume flow rate
Variable 4 Volume total
Variable 5 Density
Variable 6 Temperature
Variable 7 Drive gain
Variable 8–15 None
Alarm menu Enabled
Offline password 1234
Auto scroll Disabled
Auto scroll rate 10 sec
Table A-1 Transmitter default values and ranges continued
Type Setting Default Range Comments