Emerson 2200S Pacemaker User Manual


 
44 Micro Motion
®
Model 2200S Transmitters
Configuring Process Measurement
4.7 Configuring density measurement
The density measurement parameters control how the flowmeter measures and reports density. You
must configure:
Density measurement units
Density cutoff
Density damping
Slug flow parameters
Note: If you use the display, you can configure only the density measurement unit.
4.7.1 Density measurement units
The default density measurement unit is g/cm3. See Table 4-6 for a complete list of density
measurement units.
Configuration
To configure the density measurement unit:
Using the Communicator, select
Detailed Setup > Config Fld Dev Var > Density.
Using ProLink II, click
ProLink > Configuration > Density.
Using the display, see Figure 2-13.
4.7.2 Density cutoff
If the density measurement drops below the configured density cutoff, the transmitter will report a
value of 0 for density.
Table 4-6 Density measurement units
Density unit
Unit descriptionDisplay Communicator ProLink II
SGU SGU SGU Specific gravity unit (not temperature-
corrected)
G/CM3 g/Cucm g/cm3 Grams per cubic centimeter
G/L g/L g/l Grams per liter
G/ML g/mL g/ml Grams per milliliter
KG/L kg/L kg/l Kilograms per liter
KG/M3 kg/Cum kg/m3 Kilograms per cubic meter
LB/GAL lb/gal lbs/Usgal Pounds per U.S. gallon
LB/CUF lb/Cuft lbs/ft3 Pounds per cubic foot
LB/CUI lb/CuIn lbs/in3 Pounds per cubic inch
ST/CUY STon/Cuyd sT/yd3 Short ton per cubic yard