Emerson 2200S Pacemaker User Manual


 
112 Micro Motion
®
Model 2200S Transmitters
Troubleshooting
Inaccurate density reading Problem with process fluid • Use standard procedures to check
quality of process fluid.
Bad density calibration factors Verify characterization. See
Section 4.2.
Wiring problem Check the sensor circuitry. See
Section 10.24.
Bad flowmeter grounding See Section 10.14.2.
Slug flow • See Section 10.17.
Sensor cross-talk • Check environment for sensor with
similar (±0.5 Hz) tube frequency.
Plugged flow tube Check drive gain and tube frequency.
Purge the flow tubes.
Incorrect sensor orientation Sensor orientation must be
appropriate to process fluid. See the
installation manual for your sensor.
RTD failure • Check for alarm conditions and follow
troubleshooting procedure for
indicated alarm.
Physical characteristics of sensor have
changed
• Check for corrosion, erosion, or tube
damage.
Temperature reading significantly
different from process temperature
RTD failure • Check for alarm conditions and follow
troubleshooting procedure for
indicated alarm.
Temperature reading slightly different
from process temperature
Sensor leaking heat • Insulate the sensor.
Unusually high density reading Plugged flow tube Check drive gain and tube frequency.
Purge the flow tubes.
Incorrect K2 value • Verify characterization. See
Section 4.2.
Unusually low density reading Slug flow See Section 10.17.
Incorrect K2 value • Verify characterization. See
Section 4.2.
Unusually high tube frequency Sensor erosion • Contact Micro Motion.
Unusually low tube frequency Plugged flow tube, corrosion, or erosion Purge the flow tubes.
Unusually low pickoff voltages Several possible causes • See Section 10.23.3
.
Unusually high drive gain Several possible causes See Section 10.23.2.
Table 10-4 Process variables problems and remedies continued
Symptom Cause Suggested remedy