Erratic non-zero flow rate under
no-flow conditions
Leaking valve or seal • Check pipeline.
Slug flow • See Section 10.17.
Plugged flow tube Check drive gain and tube frequency.
Purge the flow tubes.
Incorrect sensor orientation Sensor orientation must be
appropriate to process fluid. See the
installation manual for your sensor.
Wiring problem Check the sensor circuitry. See
Section 10.24.
Vibration in pipeline at rate close to
sensor tube frequency
• Check environment and remove
source of vibration.
Damping value too low • Check configuration. See
Section 4.10.
Mounting stress on sensor Check sensor mounting. Ensure:
- Sensor is not being used to support
pipe.
- Sensor is not being used to correct
pipe misalignment.
- Sensor is not too heavy for pipe.
Sensor cross-talk • Check environment for sensor with
similar (±0.5 Hz) tube frequency.
Erratic non-zero flow rate when flow
is steady
Slug flow • See Section 10.17.
Damping value too low • Check configuration. See
Section 4.10.
Plugged flow tube Check drive gain and tube frequency.
Purge the flow tubes.
Excessive or erratic drive gain • See Section 10.23.2
Output wiring problem Verify wiring between transmitter and
receiving device. See the installation
manual for your transmitter.
Problem with receiving device • Test with another receiving device.
Wiring problem Check the sensor circuitry. See
Section 10.24.
Inaccurate flow rate or batch total Bad flow calibration factor • Verify characterization. See
Section 4.2.
Inappropriate measurement unit • Check configuration. See
Section 10.20.
Bad sensor zero Rezero the flowmeter. See
Section 7.2.
Bad density calibration factors Verify characterization. See
Section 4.2.
Bad flowmeter grounding See Section 10.14.2.
Slug flow • See Section 10.17.
Problem with receiving device • See Section 10.16.
Wiring problem Check the sensor circuitry. See
Section 10.24.