36 Micro Motion
®
Model 2200S Transmitters
Configuring Process Measurement
Configuration
To configure flow direction:
Using the Communicator, select
Detailed Setup > Config Fld Dev Var > Flow.
Using ProLink II, click
ProLink > Configuration > Flow.
Note: You cannot configure flow direction with the display.
4.3.2 Flow damping
Before configuring flow damping, review the information in Section 4.10.
Configuration
To configure flow damping:
Using the Communicator, select
Detailed Setup > Config Fld Dev Var > Flow.
Using ProLink II, click
ProLink > Configuration > Flow.
Note: You cannot configure flow damping with the display.
Table 4-2 Effect of flow direction on totalizers and digital communications
Flow direction value
Forward flow
Flow totals Flow values via digital comm.
Forward only Increase Positive
Reverse only No change Positive
Bidirectional Increase Positive
Absolute value Increase Positive
(1)
(1) Refer to the digital communications status bits for an indication of whether flow is positive or negative.
Negate/Forward No change Negative
Negate/Bidirectional Decrease Negative
Flow direction value
Zero flow
Flow totals Flow values via digital comm.
All No change 0
Flow direction value
Reverse flow
Flow totals Flow values via digital comm.
Forward only No change Negative
Reverse only Increase Negative
Bidirectional Decrease Negative
Absolute value Increase Positive
(1)
Negate/Forward Increase Positive
Negate/Bidirectional Increase Positive