3 - 56
3.4 Buffer Memory
3.4.21 CH[ ] Process alarm upper/lower limit values (Un\G94 to Un\G125)
3
SPECIFICATIONS
3.4.21 CH[ ] Process alarm upper/lower limit values (Un\G94 to Un\G125)
Process alarm upper upper limit value, upper lower limit value, lower upper limit value, and
lower lower limit value can be set. (Refer to Section 3.2.4)
(1) Setting methods
(a) A scaling range of measured temperature values is set for each channel in units of
0.1 .
Settable range and default value differ according to the thermocouple type.
The following table shows settable range and default value for each thermocouple
type.
(b) Turn on Operating condition setting request (Y9) to activate the setting.
POINT
(1) If any of the following values are set, an error (error code: 6 ) occurs.
Then, Error flag (XF) turns on.
• A value out of the above settable range:
• A value that does not satisfy the following condition:
Process alarm lower lower limit value lower upper limit value upper lower limit value
upper upper limit value
(2) When “Disable” (“0”) is set in the Warning output enable/disable setting
(Un\G46), the setting of CH Process alarm upper/lower limit values
(Un\G94 to Un\G125) is ignored.
Table 3.16 Process alarm settable range and default value
Thermocouple
type
Default value
Settable temperature range (Accuracy
gurantee range)
Process
alarm
lower
lower
limit
value
Process
alarm
lower
upper
limit
value
Process
alarm
upper
lower
limit
value
Process
alarm
upper
upper
limit
value
Thermocouple K -2000 12000 -2700 to 13700 (-2000 to 12000)
Thermocouple E -2000 9000 -2700 to10000 (-2000 to 9000)
Thermocouple J -400 7500 -2100 to12000 (-400 to 7500)
Thermocouple T -2000 3500 -2700 to 4000 (-2000 to 3500)
Thermocouple B 6000 17000 0 to18200 (6000 to17000)
Thermocouple R 0 16000 -500 to17600 (0 to16000)
Thermocouple S 0 16000 -500 to17600 (0 to16000)
Thermocouple N -2000 12500 -2700 to13000 (-2000 to12500)