Mitsubishi Q68TD-G-H02 Thermometer User Manual


 
3
SPECIFICATIONS
3.2 Function List
3.2.2 Conversion setting for disconnection detection function (Q68TD-G-H02 only)
3 - 13
1
OVERVIEW
2
SYSTEM
CONFIGURATION
3
SPECIFICATIONS
4
SETUP AND
PROCEDURES BEFORE
OPERATION
5
UTILITY PACKAGE (GX
CONFIGURATOR-TI)
6
PROGRAMMING
7
ONLINE MODULE
CHANGE
8
TROUBLESHOOTING
(2) Setting methods
When “Up scale”, “Down scale” or “Given value” is selected in Conversion setting for
disconnection detection (Un\G164, Un\G165), the value to be stored into CH
Measured temperature value (Un\G11 to Un\G18) in disconnection detection can be
specified.
(a) Up scale and down scale
When “Up scale” (0H) or “Down scale” (1H) is selected, an up scale value or down
scale value of thermocouple-type to be used is stored into the CH Measured
temperature value (Un\G11 to Un\G18).
(b) Given value
When “Given value” (2H) is selected, set a value in the CH Conversion setting
value for disconnection detection (Un\G166 to Un\G173) in units of 0.1 .
The value set in the buffer memory above is stored into the CH Measured
temperature value (Un\G11 to Un\G18) when disconnection is detected.
Table 3.4 Upscale and down scale list
Thermocouple
type
Setting value
Measurement
range
Measured temperature value when
disconnection is detected.
Up scale Down scale
Thermocouple K 0
-270 to 1370 1452.0 -352.0
Thermocouple E 1
-270 to 1000 1063.5 -333.5
Thermocouple J 2
-210 to 1200 1270.5 -280.5
Thermocouple T 3
-270 to 400 433.5 -303.5
Thermocouple B 4
0 to 1820 1911.0 -91.0
Thermocouple R 5
-50 to 1760 1850.5 -140.5
Thermocouple S 6
-50 to 1760 1850.5 -140.5
Thermocouple N 7
-270 to 1300 1378.5 -348.5