5-1 OVERVIEW
Ensure that the oxygen analyzer, heater power
supply, and intelligent field transmitter have
been properly connected. It is important to
check that grounding and screening of termina-
tions are correctly made to prevent the introduc-
tion of ground loops. The IFT is equipped with
noise suppression circuitry on the power supply
and signal input lines. Proper grounding at in-
stallation will ensure accuracy of function.
The following five languages are can be se-
lected within the IFT:
English Italian
French Spanish
German
NOTE
Support the keypad with the free hand
to prevent bounce back of the IFT
door.
a. Intelligent Field Transmitter (IFT)
The Intelligent Field Transmitter may be
supplied with either of two configurations.
These are the blind version and the deluxe
version. The two versions differ as follows:
1. Blind Version. The blind version has
no display and no keypad. With this
version an external HART communica-
tions device is required.
2. Deluxe Version (GUI). The deluxe
version is also known as the General
User Interface (GUI) version. This IFT
contains an LED display, liquid crystal
display panel, and an eight-key pad
from which the probe and electronics
can be configured, calibrated and
troubleshooted.
b. HART Communicator Interface Devices
The HART communications protocol can
interface with any of the above IFT ver-
sions. To interface a HART communicator
with an IFT, one of three interface devices
is required. The interface devices are as
follows:
1. Rosemount Analytical Model 275/375
Handheld Communicator. The hand-
held communicator requires Device
Descriptor (DD) software specific to the
World Class 3000 product. The DD
software will be supplied with many
model 275/375 units, but can also be
programmed into existing units at most
Fisher-Rosemount service offices.
2. Personal Computer (PC). The use of a
personal computer requires Corner-
stone software with Module Library
(ModLib) specific to the World Class
3000 product.
3. Selected Distributed Control Systems.
The use of distributed control systems
requires input/output (I/O) hardware
and software which permit HART
communications.
This section of the manual deals with op-
erator controls and displays available with
the GUI equipped IFT. Operating parame-
ters are listed and instructions are included
for viewing and changing them.
Any procedures not associated with normal
operation are included in Section 2, Instal-
lation, or Section 5, Troubleshooting.