Emerson 3000 Oxygen Equipment User Manual


 
Instruction Manual
IB-106-300NH Rev. 4.3
May 2005
2-20 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
ALWAYS DISCONNECT LINE VOLTAGE
FROM HEATER POWER SUPPLY AND
ANALOG ELECTRONICS (IF USED)
BEFORE CHANGING JUMPERS.
JUMPER
CONFIGURATIONS
LINE VOLTAGE
SELECTION
JUMPER
(INSTALL)
JUMPER
(INSTALL)
HEATER
POWER
JUMPER
100/120 VAC
220/240 VAC
JM4, JM1
JM5
REMOTE
ON
REMOVE JM2
*INSTALL JM2
ELECTRONICS
SELECTION
JUMPER
PROBE HEATER
VOLTAGE SELECTION
INSTALL JM3, JM6
REMOVE JM3, JM6
*ANALOG (EXISTING)
DIGITAL
(NEXT GENERATION)
WORLD CLASS PROBE
(44V)
JM7
JM8
218 PROBE (115V)
1
2
NOTES:
100 VAC OPERATION REQUIRES TRANSFORMER PART
NUMBER 1M02961G02.
REFER TO TABLE 3-5 FOR PROPER SET POINT SELECTION.
1
2
37840009
Figure 2-15. Jumper Selection Label
NOTE
Before supplying power to the heater
power supply, verify that jumpers JM3,
JM6 are removed and JM7 is installed.
If relay wire (Figure 2-13, note 1) is in-
stalled, JM2 must be removed from
HPS Mother Board (Figure 2-16).
4. Before supplying power to the heater
power supply, verify that the jumpers
on the mother board, Figure 2-16, are
properly configured. Jumpers JM3 and
JM6 should be removed and JM7
should be installed.
Additionally, make sure that the proper
jumper for your line voltage is installed,
Figure 2-15. If relay wire (Figure 2-13,
note 1) is not installed, JM 2 should be
installed on the HPS Mother Board
(Figure 2-16).
NOTE
Refer to Figure 2-8 and Figure 2-10 for
proper IFT jumper configuration. IFT
microprocessor and interconnect
board jumper configurations must be
set correctly in order for HPS to work
properly.
29850001
NOTE: ALL FUSES ARE 5A @ 250 VAC,
ANTI-SURGE, CASE SIZE
5 X 20 MM, TYPE T TO IEC127,
SCHURTER.
3D3 080G REV
JM1
JM2
JM4
JM5
JM7
FUSE
FUSES
Figure 2-16. Jumpers on HPS Mother Board