2-12 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
BK
GN
WH
PU
SHIELD
SHIELD
CAL RET
NO GAS
LO GAS
HI GAS
IN CAL
SHIELD
SHIELD
STACK TC -
STACK TC +
PROBE
M
V
+
PROBE MV
-
PROBE TC +
PROBE TC -
RD
YE
BL
OR
BL
OR
YE
RD
INTELLIGENT FIELD TRANSMITTER IFT 3000
WORLD CLASS
PROBE
12345678
BL
PROBE MV -
PROBE MV +
PROBE TC +
PROBE TC -
OR
YE
RD
GN
E
R
H
WH
BK
GN CELL -VE
OR CELL +VE
YE CHROMEL
RD ALUMEL
GN
BK
BK
}
HEATER
PROBE JUNCTION
BOX WIRING
STACK TC WIRING AS REQUIRED.
SPECIAL PROBE CABLE BETWEEN PROBE
AND IFT BY ROSEMOUNT ANALYTICAL.
INSTALL JM1 ON INTERCONNECT BOARD.
INSTALL JM6 ON MICROPROCESSOR
BOARD PER FIGURE 2-8.
IF STACK TEMPERATURE NOT USED.
IF MPS 3000 NOT USED.
1 RELAY PER PROBE AVAILABLE FOR
CALIBRATION STATUS INDICATION. (48 V
max, 100 mA max)
CURRENT/VOLTAGE SELECTOR SWITCH
MUST BE SELECTED TO CURRENT (I) FOR
HART COMMUNICATIONS APPLICATIONS.
SET SWITCH SW3A PER FIGURE 2-8.
SWITCH/JUMPER CONFIGURATION FOR REV.
13 AND LATER
NOTES:
ALWAYS DISCONNECT LINE VOLTAGE
FROM INTELLIGENT FIELD TRANSMITTER
BEFORE CHANGING JUMPERS.
LINE
VOLTAGE
SECTION
JUMPER
(INSTALL)
PROBE HEATER
VOLTAGE SECTION
JUMPER
(INSTALL)
120V.A.C.
100V.A.C.
220V.A.C.
240V.A.C.
J1
J5 J6
J1
JM1
J2
J3
J4
J5
J6
J7
J8
J9
3D39122GREV
POWERSUPPLYBOARD
3D39513G,REV13
MICROPROCESSOR
BOARD
3D39120GREV
INTERCONNECTBOARD
JM8,JM7,JM1
JM3,JM7,JM2
JM6,JM5,JM2
JM6,JM5,JM1
WORLDCLASSPROBE
WORLDCLASS"DIRECT
REPLACEMENT"PROBE
218PROBE
JM10
JM9
JM9
JUMPERCONFIGURATION
5 CONDUCTOR SHIELDED CABLE
PER PROBE #16 AWG BY CUSTOMER
EN
L
LINE
VOLTAGE
MPS TERMINATION BOARD
MPS 3000 MULTIPROBE CALIBRATION GAS SEQUENCER (OPTIONAL)
ER
H
L
E
N
LINE
VOLTAGE
J13 J14 J15 J16 J17 J18
J12
CAL RET
HI GAS
IN CAL
NO GAS
CAL RET
HI GAS
IN CAL
NO GAS
CAL RET
HI GAS
IN CAL
NO GAS
CAL RET
HI GAS
IN CAL
NO GAS
LOW GAS
LOW GAS
LOW GAS
LOW GAS
NC C NO NC C NO NC C NO NC C NO
L
N
L
N
LINE OUT LINE IN
J10
J11
PROBE 1
PROBE 2 PROBE 3 PROBE 4
PROBE 1
PROBE 2 PROBE 3 PROBE 4
PROBE 1
SOLENOID
PROBE 2
SOLENOID
PROBE 3
SOLENOID
PROBE 4
SOLENOID
HIGH GAS
SOLENOID
LOW
GAS
SOLENOID
PRESSURE
SWITCH
CURRENT/VOLTAGE
SELECTOR SWITCH
GND
STUD
AOUT -
AOUT +
2 CONDUCTOR
SHIELDED CABLE
BY CUSTOMER
10
J1
JM6
J2
J3
SW2
SW3
10
A
B
C
D
ON OFF
37840006
Figure 2-7. Wiring Layout for IFT Systems without HPS