World Class 3000
LEGEND FOR FIGURE D-6
1. Enclosure
2. Screw
3. Plug
4. Cable Grip
5. Fitting
6. Hose Adapter
7. Pressure Switch
8. Plug
9. Solenoid Valve
10. Screw
11. Manifold
12. Washer
13. Screw
14. Gasket
15. Outer Cover
16. Inner Cover
17. Flowmeter, 10 SCFH
18. Flowmeter, 2.0 SCFH
19. Bracket
20. Screw
21. Hose Adapter
22. 1/8 in. Hose
23. Nut
24. Lockwasher
25. Washer
26. Washer
27. Screw
28. Nut
29. Washer
30. Washer
31. Cover Stop Slide
32. Screw
33. Washer
34. Termination Board
35. Standoff
36. Mounting Bracket
37. Cover Plate
38. Washer
39. Screw
40. Fuseholder
41. Plastic Nut
42. Bushing
43. Pressure Gauge
44. Bolt
45. Washer
46. Drain Valve
47. 1/8 in. Impolene Tubing
48. Connector
49. Elbow
50. Pressure Regulator
51. Hose Adapter
52. 1/4 in. Tube
53. Screw
54. Washer
55. Inner Enclosure
56. Washer
57. Screw
58. Power Supply
59. Washer
60. Nut
61. Screw
62. Washer
63. Solenoid
64. Solenoid
e. Loosen the retaining ring in the middle of
the solenoid and remove the top part.
f. With a spanner wrench or padded pliers,
remove the remaining part of the solenoid
from the manifold (11).
g. Separate the new solenoid and screw the
smaller part into the manifold.
h. Place the top part of the solenoid into posi-
tion and tighten the retaining ring.
i. Connect the plug to the proper receptacle
on the termination board (34).
j. Close and secure the inner cover (16) with
two captive screws. Close and secure the
outer cover (15) with two captive screws.
D-9 PRESSURE REGULATOR MAINTENANCE
a. Pressure Adjustments
Pressure regulator (50, Figure D-6) is fac-
tory set to 20 psi (138 kPa). Should the
pressure need to be changed or adjusted,
use the knob on top of the pressure
regulator.
b. Condensation Drain
To drain excess moisture from the internal
gas circuit of the MPS, periodically loosen
drain valve (46) on the bottom of pressure
regulator (50). The moisture will flow
through vinyl tubing drain (47) on the bottom
of pressure regulator (50) and exit the bot-
tom of MPS enclosure (1).