Emerson 3000 Oxygen Equipment User Manual


 
Instruction Bulletin
Appendix E Rev. 4.6
May 2005
E-6 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
YES
YES
OXYGEN ANALYZER
SYSTEM IS EQUIPPED
WITH HPS 3000.
CHECK JUMPERS JM6 ON
MICROPROCESSOR BOARD
ANDJM1 ON INTERCONNECT
BOARD ARE INSTALLED.
CHECK JUMPERS9
OR 10ON POWER
SUPPLY BOARD FOR COR-
RECT CONFIGURATIONS.
REFER TO PROBE
T
ROUBLESHOOTING,
APPENDIX A, SECTION 2.
CHECK THAT JUMPERS
JM6 ONMICROPROCESSOR
BOARD ANDJM1 ON
INTERCONNECTBOARD
ARE REMOVED.
REFER TO APPENDIX B,
SECTION 2,FOR
JUMPER LOCATIONS.
DISCONNECT POWER
AND INSTALL JU
MPERS
CORRECTLY. REFER TO
FIGURE 2-5OF MAIN IB
SECTION.
NO
NO
SYMPTOM - MICROPROCESSOR BOARD LED ISSTEADY ON
27270021
Figure E-4. IFT Troubleshooting Flowchart, #2
END OF FLOWCHART #2
SYMPTOM - GENERAL USER INTERFACE OR
LED DISPLAY PANEL DOES NOTFUNCTION
REPLACE GUIBOARD.
STILL NOTFUNCTIONING.
CALL FACTORYFOR
ASSISTANCE.
27270020
Figure E-5. IFT Troubleshooting Flowchart, #3